Skip Navigation


QJM Advance Access originally published online on March 9, 2009
QJM 2009 102(5):335-339; doi:10.1093/qjmed/hcp020
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
102/5/335    most recent
hcp020v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Cooke, J.
Right arrow Articles by Lyons, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Cooke, J.
Right arrow Articles by Lyons, D.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author 2009. Published by Oxford University Press on behalf of the Association of Physicians. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org

Sitting and standing blood pressure measurements are not accurate for the diagnosis of orthostatic hypotension

J. Cooke, S. Carew, M. O’Connor, A. Costelloe, T. Sheehy and D. Lyons

From the Clinical Age Assessment Unit, Mid-Western Regional Hospital, Dooradoyle, Limerick, Ireland

Address correspondence to Dr J. Cooke, Clinical Age Assessment Unit, Mid-Western Regional Hospital, Dooradoyle, Limerick, Ireland. email: john.cooke{at}hse.ie

Received 11 November 2008 and in revised form 20 January 2009


   Abstract

Introduction: Orthostatic hypotension (OH) is associated with troublesome symptoms and increased mortality. It is treatable and deserving of accurate diagnosis. This can be time consuming. The current reference standard for its diagnosis is head-up tilt (HUT) testing with continuous beat-to-beat plethysmography. Our objective was to assess the accuracy of sit-stand testing with semi-automatic sphygmomanometry for the diagnosis of OH.

Design: Retrospective test of diagnostic accuracy.

Methods: This was a retrospective study performed using a database maintained by a busy syncope unit. HUT testing was performed using an automated tilt table with Finometer monitoring. A 3 min 70° HUT was performed following 5 min supine. Sitting blood pressure (BP) was measured following 3 min rest. Standing BP was measured within 30 s of assuming the upright posture. The results of sit-stand testing were compared with HUT testing as a reference standard. Both tests happened within 5 min of each other and patients underwent no intervention between tests.

Results: From a total of 1452 consecutive HUTs, we identified 730 with pre-test measures of sitting and standing BP. The mean age of this group was 70.57 years (SD = 15.1), 62% were female. The sensitivity of sit-stand testing was calculated as 15.5%, specificity as 89.9%, positive predictive value as 61.7%, negative predictive value as 50.2% and the likelihood ratio as 1.6. The area under the Receiver Operator Curve was 0.564.

Conclusion: We have demonstrated that sit-stand testing for OH has very low diagnostic accuracy. We recommend that the more time-consuming reference standard method of diagnosis be used if the condition is suspected.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.